4.1Version 2.2.0
- At least 6 samples will be ignored after range changed no matter if up or down
- Added command to temporary change the count of ignored samples after range changed (RCISC)
- Each range switch will cause a peak in min/max value (and a small error in the average value).
- This command can be used to minimize that behavior
- Added command to temporary change the minimum range (MINRG)
- Each range switch will cause a peak in min/max value (and a small error in the average value).
- This command can be used to minimize that behavior by reducing count of range switches
4.2Version 2.1.3
- Some samples will be ignored after range changed (Up: 4 Samples; Down: 6 Samples)
- Update to support an alternative operational amplifier (original one was out of stock)
- Update to support a controller type with less memory
4.3Version 1.3.0
- Added command for averaging min./max. value (MMASC)
- Added command to save user defined text (UTEXT)
4.4Version 1.2.7
- Added command to query the serial number (SERNB)
- Added command to query the calibration date (CDATE)