A temperature sensor (TS) module has the following mandatory standard parameters:
| Parameter | Meaning | Examples |
|---|---|---|
| Resource | The connection information needed to connect to this device. Depending on the type of device this can be a COM port, a GPIB address, an IP address or something else. Most drivers are based on VISA, which brings its own syntax to provide a GPIB or IP address. The VISA resource names can usually be looked up in NI Max. | COM1 |
| MinTemp | The lower temperature limit in °C. This value can be changed through the LTT GUI. The GUI will stop all running tests if this limit will be violated. | -41 |
| MaxTemp | The upper temperature limit in °C. This value can be changed through the LTT GUI. The GUI will stop all running tests if this limit will be violated. | 120 |
Depending on the specific TS module, there might be additional parameters, which are described here.
You can access the TS control panel through GUI and also via TestStand..
Driver Specific Parameters
Some temperature sensor drivers have additional parameters besides the standard ones described in here.
1.1.2.2.3.1.1PMA280
| Parameter | Meaning | Examples |
|---|---|---|
| Baud Rate | An optional parameter providing the baud rate for the PMA280 module. If not provided 9600 is the default baud rate. | 9600 19200 |
| Timeout [ms] | An optional parameter specifying the communication timeout for the PMA280 module. There are cases where you need to increase the default communication timeout, otherwise it will lead to sporadic timeout errors. If you connect via Windows Remote Desktop it is recommended to set a timeout value of 3000. | 1000 2000 3000 |